Screening
& Diagnosis
THE
LEAD LABORATORY
Dr.
Patrick J. Parsons, and Dr. J. Julian Chisolm, Jr.
Role of the Laboratory
Laboratories
measure lead concentrations in either clinical samples such as blood
and urine or in environmental samples such as paint and dust. It is
important to distinguish between the clinical and environmental lead
laboratories and the issues that they face. Often, laboratories are
separated according to the types of samples they analyze even though
the technologies used to measure lead levels may be similar.
Several reasons
exist for separating laboratories along sample lines. For example, clinical
and environmental laboratories operate under different federal or state
regulations. Sample handling and reporting requirements of state and
federal agencies differ, as do requirements for quality control. The
reportable concentration range for lead in paint or in dust, and the
preparation of the sample requires different procedures.
Further, the higher
concentrations of lead typically found in lead-based paint (>0.5%
by weight) and in lead contaminated dust (>1 mg/g) present the potential
for contaminating the laboratory environment. If environmental and clinical
laboratory functions are carried out in the same laboratory room, determination
of the much lower concentrations of lead found in blood is jeopardized.
For all clinical
laboratories in the United States, the Health Care Financing Agency
(HCFA), operating under the Clinical Laboratory Improvement Amendments
of 1988 (CLIA), regulates all aspects of laboratory operation including
determining the qualifications of a laboratory director and establishing
protocols for quality assurance and quality control (QA/QC) activities,
method validation, specimen collection, storage, analysis and reporting
of results (42 CFR Part 493). For blood lead laboratories specifically,
successful participation in an approved proficiency testing (PT) program
is required. In addition, many states regulate the operation of clinical
laboratories under their jurisdiction although the requirements vary
greatly from state to state. Laboratories are advised to contact their
state health department to ensure compliance with state and local laws.
For environmental
laboratories, the primary regulatory agencies are the U.S. Environmental
Protection Agency (EPA) and the National Institute for Occupational
Safety and Health (NIOSH). In many laboratories, the EPA’s Contract
Laboratory Program (CLP) requirements drive all QA/QC activities and
restrict the analytical protocol used. Several states also regulate
environmental laboratories under their jurisdiction, although regulations
vary from state to state.
The clinical
lead laboratory
Determining
lead levels in blood
A blood lead test
is recommended for initial screening purposes, and for diagnostic evaluation.
Serial blood lead measurements are recommended for monitoring patients
under medical management.
Specimen collection
Capillary blood
lead measurements may be used for initial screening purposes, but only
venous blood is appropriate for diagnostic evaluation, and a venous
BLL is necessary before initiating an environmental investigation or
chelation therapy. From 1991 to 1994, CDC sponsored several studies
to evaluate the performance of capillary blood lead screening by comparing
blood lead levels in specimens obtained by fingerstick with those obtained
simultaneously by venipuncture.
Results of studies
conducted by Yale New Haven Hospital and the Connecticut Department
of Health, the City of Milwaukee, and the New York State Department
of Health, independently concluded that, provided that a strict cleaning
protocol is followed, capillary whole blood obtained by fingerstick
is a reliable specimen for pediatric lead screening purposes.
Most capillary blood
specimens that show falsely elevated lead levels can be traced to inappropriate
collection procedures or to contaminated materials used to collect and
transport the specimen1. For these reasons, laboratories
should ensure that all materials used to collect capillary blood specimens
are free from significant lead contamination. It is recommended that
capillary blood be collected in plastic microcollection devices containing
either EDTA (lavender caps) in powder form or heparin (green caps) rather
than glass capillary tubes, which can break and may result in injury
and disease transmission.
The choice of microcollection
container is a matter for each individual laboratory and depends on
factors such as the desired sample volume or anticoagulant type. Although
capillary blood is useful for initial screening, only venous blood should
be used for diagnostic evaluation or medical management. A recommended
procedure for collecting capillary blood by fingerstick is given in
Appendix C.2.
Recommended procedures
for controlling contamination
Perhaps the most
frequent source of error is contamination by airborne particulates containing
lead. Although use of lead in gasoline and paint has been reduced, lead
still remains a ubiquitous pollutant, especially in common dust. Therefore,
special precautions must be taken to prevent contamination during specimen
collection and analysis. These precautions include checking all collection
materials and supplies for significant contamination and using dust-control
measures in the laboratory as recommended below.
Some manufacturers
provide blood collection tubes specifically for trace element analysis
(royal blue caps) or certified lead-free tubes for blood lead level
determination (tan/brown caps). Laboratories should check lead-free
claims made by individual manufacturers before using such devices since
the amount of lead in these devices is reported to vary greatly.
Once a reliable
source of certified lead-free materials has been located, laboratories
need not check each individual lot for lead contamination. Since the
cost of certified lead-free tubes is likely to be much more than for
standard tubes, some laboratories may prefer to obtain the standard
tubes and certify them as lead-free for their clients as described below.
This strategy will hold down the costs associated with blood lead level
(BLL) screening.
Checking collection
materials and supplies
Two approaches may
be used to assess contamination and ensure that materials are not significantly
contaminated with lead. First select at least 10 collection devices
at random from a batch to test. Then take either of these two steps:
- Fill specimen
or sample containers with either dilute acetic acid (4% v/v) or dilute
nitric acid (2% v/v) and store for 24 hours at room temperature. Analyze
leachate for lead and alculate the total amount of lead extracted.
- Fill specimen
or sample container with a base low-lead blood sample of known lead
concentration, and store for a period and under conditions the the laboratory
maintains for routine patient samples. Analyze the blood lead levels
to ascertain whether or not any significant lead contamination has occurred.
For both methods,
any measurable increase in lead concentration should amount to no more
than 0.5 mg/dL (i.e., 5% at 10 mg/dL). Needles, lancets, and materials
other than containers should also be checked for gross contamination
by leaching with a minimum volume of dilute acetic or dilute nitric
acid and analyzing the leachate for lead. Generally, such materials
should be free of significant contamination (i.e. <1 mg/L or roughly
the detection limit by graphite furnace atomic absorption spectrometry
[GFAAS]).
In the laboratory,
the handling of all blood is governed by the Occupational Safety and
Health Administration’s (OSHA’s) blood borne pathogens rule (29 CFR
Part 1910.1030), which specifies that CDC Universal Precautions must
be used. For example, blood dilutions may be performed in Class II biosafety
cabinet using powder-free protective gloves. This procedure will not
only protect the analyst from pathogens but will also protect the sample
from airborne lead contamination.
Most clinical laboratories
are not equipped with Class 100 air or better, nor do they need to be,
as long as some basic consideration is given to the potential for airborne
contamination at the bench. Autosampler vials should be screened for
lead contamination until a lead-free supply is verified and should be
stored protected from dust (e.g., in sealed plastic bags). Autosampler
vials that are in use should be protected with appropriate dust covers.
Frequent wet-mopping of laboratory floors and wet-wiping of other flat
surfaces will minimize contamination from airborne dust particulates.
Transporting
or Shipping Blood Specimens to Centralized Laboratories
Currently, most
blood lead testing is performed by a centralized laboratory that is
licensed for this purpose. Consideration needs to be given to the most
cost-effective and reliable way to transport specimens to the laboratory.
Although some laboratories may use a network of couriers to transport
specimens to the laboratory, others rely on the United States Postal
Service (USPS) or commercial carriers to deliver specimens. Transport
of etiologic agents (e.g., human blood) via USPS is regulated (42 CFR
Part 72), and many commercial carriers may have similar restrictions.
Federal regulations
require that blood specimens be packaged according to guidelines requiring
the use of tertiary containment and sufficient absorbent material in
the event that the specimen leaks during transit. When mailing glass
tubes containing blood, ensuring that the tubes cannot touch during
transit may reduce chances of breakage.
Venous blood specimens
preserved with EDTA or heparin are reportedly stable for determining
lead levels in blood for up to 10 weeks if they are refrigerated at
4°C2. Refrigerated temperatures are not necessary for mailing
blood lead specimens. However, where significant delays are expected,
as might occur over a holiday weekend, it would be prudent to store
the samples locally, refrigerated at 4°C, and then ship them the next
business day. In special circumstances, such as stat requests,
blood specimens should be mailed overnight and the laboratory forewarned
to expect them.
Analytical Methods
for Determining Blood Lead Levels
Several analytical
methods have been applied successfully to the determination of lead
in blood. The methods most commonly in use today are either GFAAS or
anodic stripping voltammetry (ASV). Other analytical methods include
variations on AAS, such as methylisobutylketone (MIBK)-extraction flame
AAS,3 Delves-cup microsampling flame AAS,4 and
an inductively-coupled plasma interfaced to a quadruple mass spectrometer
(ICP-MS).
Beginning in 1992,
CDC funded several innovative research projects aimed at developing
portable instrumentation for determining lead levels in blood. Techniques
under investigation include attempts to miniaturize AAS and atomic emission
spectrometry (AES) and to develop various electrochemical approaches
and polymeric sensors. Although each of these techniques has advantages
and disadvantages, and although the new technologies may result in commercially
available instruments in the future, we review only GFAAS and ASV in
detail here.
Graphite Furnace
Atomic Absorption Spectrometry
GFAAS has been successfully
used to determine lead levels in blood.5-8 Modern furnace
instrumentation is reliable, accurate and precise and can be reasonably
automated. Several manufacturers currently market GFAAS instruments
that are readily configured for blood lead testing. Most of these instruments
are also capable of measuring many more elements.
Selecting suitable
GFAAS instruments should be done after a careful evaluation of available
commercial instruments. Prospective users should request that the manufacturers
provide references from customers currently performing blood lead determinations
and provide evidence of successful participation in an approved PT program
for measuring blood lead levels. Numerous instrumental configurations
and features are available.
Many GFAAS methods
for measuring blood lead levels using a variety of furnace instruments9
have been published.* Until recently, it was thought that no one method
for determining blood lead levels could be recommended for all instrumentation
because of the complexity and differences between furnace equipment
from the various manufacturers. However, recent experience has shown
that a common or standard method for determining blood lead by GFAAS
is possible.8 This method, which uses longitudinal Zeeman
background correction and a transversely-heated furnace draws upon the
work of both Pruszkowska-et al.,5 and Miller et al.7
This simple, but rapid Zeeman AAS method for measuring blood lead levels
has been successfully transferred to a simpler furnace arrangement that
uses continuum background correction10 and to other furnace
AAS equipment using transverse Zeeman, continuum, and Smith-Heifjte
background correction systems11,12.
Such standard methods
are successful because they follow the stabilized temperature platform
furnace (STPF) concept.13 For a detailed description of the
STPF approach to GFAAS method development, consult the appropriate references.13,14
One critical component of the STPF approach is use of a suitable matrix
modifier, which is required to stabilize lead during pyrolysis at temperatures
exceeding 600°C and which will increase the volatility of the interfering
matrix. Ammonium phosphate, either in the monobasic, NH4H2PO4,
or dibasic form of ammonium phosphate, (NH4)2HPO4,
along with Triton X-100 is now the most widely used modifier for determining
blood lead levels by GFAAS.
Another important
aspect of the STPF approach calls for using the L’vov platform for atomization
coupled with integrated peak areas for absorbance measurements. Optimizing
lead atomization from the platform rather than from the wall reduces
gas-phase interferences and permits calibration against aqueous lead
standards. Precision is also much improved over peak height measurements.
Most modern GFAAS
instruments are equipped with auto samplers, which are useful for large
numbers of samples and provide for unattended operation with better
precision. The number of replicate firings that should occur depends
on the method within-run precision, which should be less than +
0.5 µg/dL at 10 µg/dL (5%). Replicate firings are recommended if the
within-run precision is greater than 10%. When using the Delves-cup
microsampling flame AAS method, where the within-run precision at 10
µg/dL can be as much as 40% (+ 4 µg/dL), carry out the analysis
in triplicate. Note that this precision is different from the concept
of analytical accuracy, for which federal and some state authorities
require results of performance samples to be accurate to within +
4 µg/dL (or + 10%, whichever is greater) of the established target
value.
Other spectrometric
techniques have been used to determine blood lead levels with varying
degrees of success. Inductively coupled plasmamass spectrometry (ICP-MS)
is a powerful technique that not only can detect very low concentrations
of lead but can also identify and quantify the lead isotopes present.
The use of ICP-MS to determine blood lead levels includes limited attempts
at source identification through isotope ratio fingerprints.15
Unfortunately, this technique is currently an expensive solution that
is not cost effective for routine blood lead measurements and is limited
to a handful of specialized research laboratories.
Flame AAS, with
chelation of lead and its extraction into MIBK solvent, is an older
technique that requires a relatively large volume (5-7 mL) of venous
blood for analysis.3 Delves-cup microsampling flame AAS was
developed specifically for determining blood lead levels in small blood
volumes,4 but, since the equipment and supplies are no longer
available from AAS instrument manufacturers, it has been largely replaced
by GFAAS.
Anodic Stripping
Voltammetry (ASV)
Anodic stripping
voltammetry (ASV) is a manual electrochemical method capable of measuring
many metals in a variety of matrices. ASV has been used to determine
blood lead levels for more than 25 years. Several versions of instruments
that are capable of performing this type of analysis are commercially
available. The ASV technique is straightforward, and the instrumentation
is relatively inexpensive, but reagent cost per test is higher than
for GFAAS. ASV instruments are small, occupying less bench space than
GFAAS, and require no special facilities. (GFAAS generally requires
a 220 VAC, 30 A power supply.)
The analytical principle
requires that lead be decomplexed and available for plating as the free
+2 aqueous cation (Pb2+aq). One approach is to
digest (wet ash) blood specimens with nitric acid, a time-consuming
and laborious preanalytical step. More commonly, blood specimens are
incubated in a solution containing several other metal ions that compete
with lead for typical binding sites (proteins, amino acids, EDTA, and
other biological ligands), and thus displace lead as the free cation.
One ASV instrument
manufacturer provides a proprietary decomplexation solution along with
two controls (or, more appropriately, calibrators) specifically for
blood lead analysis. To perform a typical ASV analysis, a laboratorian
uses a pipette to place 100 µL of anticoagulated whole blood into a
tube containing a premeasured volume (2.9 mL) of the decomplexing reagent.
The tube is placed in contact with the instrument electrode, and the
analytical-cycle is started. The concentration of lead in the blood
is available in units of µg/dL within 90 seconds.
Most commonly employed
instruments use either the Zeeman-effect or continuum background correction
systems, and some have successfully used the Smith-Heifjte background
correction system. Each system can correct for nonspecific background
at 283.3 nm, the principal analytical wavelength for lead.
Using the Zeeman-effect
background correction system is more straightforward than using a continuum
source; although the latter is less expensive, it can be difficult to
optimize since two radiation sources (hollow cathode and continuum)
are used, both of which must be carefully aligned along the optical
path of the graphite furnace.
During the first
60 seconds of the 90-second analysis time, a negative potential is applied
to a mercury-coated graphite electrode. (Other electrode designs have
also been used.) This process results in the lead (as Pb2+)
plating-out (or dissolving) into the mercury coat. Usually, the potential
is automatically and linearly swept back in a positive anodic direction.
Other manufacturers may recommend functions other than a linear sweep
(e.g., a square wave ASV). At a specific and characteristic voltage
the lead is stripped from the electrode, a process that produces currents,
the sum of which are proportional to the concentration of lead in the
original sample. The samples may be prepared singularly or in a batch,
and the analytical cycle is repeated for each sample. The operating
parameters for analysis are normally established by the manufacturer
and are set at installation. Each day the integration window, which
must be symmetrically positioned about the lead peak, is verified and,
if necessary, reset. This parameter is referred to as the integration
set point.
Anticoagulants
for ASV
Either heparin or
EDTA may be used with the ASV method. Heparin requires no special procedures
for analysis but may present a problem because of micro-clot formation,
which makes the blood specimen non-analyzable. This problem most often
occurs when blood specimens must be transferred from the collection
site to a remote laboratory. EDTA is the preferred anticoagulant.
After the samples
have been placed in tubes using a pipette, they should be incubated
for 15-30 minutes before proceeding with the analysis. If standard evacuated
glass tubes containing EDTA are used to collect blood, then the tubes
must be at least half-full before the samples can be analyzed because,
if the concentration of EDTA exceeds the usual 1.5 mg/mL, decomplexation
of the lead is reduced, and the level of lead determined will be falsely
low.
Calibration
When direct determination
of lead levels in blood is performed by ASV (i.e., without acid digestion),
calibration with aqueous lead standards is not possible, nor is it possible
to use lead-spiked blood, since there are suitable differences between
whole blood containing endogenous lead and whole blood spiked with inorganic
lead. The instrument should be calibrated with blood-based materials
traceable to the National Institutes of Standards and Technology (NIST)
or to another well-characterized reference material. Blood-based calibrators
may also be purchased commercially, prepared by the laboratory with
the assistance of a reference laboratory, or they may be provided, along
with the reagents, by the instrument manufacturer.
One reference laboratory
with successful experience using ASV reports that calibration is most
accurate if human blood, with endogenous lead levels that have been
established by thermal ionization mass spectrometry (TI-MS), is used
(J. Chisolm, MD, Kennedy Krieger Institute, personal communication,
1993). Certainly, other ASV reference laboratories16 would
also agree that success with determining blood lead levels depends on
routinely checking instrument calibration with independently-validated
blood-based reference materials, such as the Standard Reference Material
(SRM) 955a Lead in Blood from NIST, the CDC Blood Lead Laboratory Reference
System (BLLRS) pools, or New York State’s lyophilized reference material
for blood lead determinations.
Another concern
for ASV users is a potential interference from copper, which is oxidized
or stripped at a potential close to that for lead.16 Since
copper is an essential element and is present in human serum in detectable
amounts, the possibility exists for an interference, especially in populations
where serum copper levels are elevated (e.g., in pregnant women). This
interference is minimized by properly selecting the integration set
point and using the currently recommended instrument parameters. Some
laboratories use a strip-chart recorder to check for problems with copper
interference.
Other Electrochemical
Methods for Measuring Blood Lead Levels
Recent developments
in analytical instrumentation have led to the development of prototype
instruments for determining lead levels in blood by potentiometric stripping
analysis (PSA), an electrochemical technique similar to ASV. Although
not yet commercially promoted for measuring blood lead levels, PSA has
been used successfully.17,18 In addition, other electrochemical
technologies are currently under development with the goal of providing
a relatively inexpensive and portable means of screening children for
lead poisoning.
Quality Assurance
and Quality Control in the Clinical Lead Laboratory
Quality assurance
and quality control can be loosely defined as those aspects of laboratory
policy and practice which ensure that all test results are reported
accurately. Although it is not possible to give a comprehensive description
of all QA/QC practices, some aspects that are considered desirable for
the clinical lead laboratory are given below. Under the 1988 Clinical
Laboratory Improvement Act (CLIA) regulations, all clinical laboratories
in the United States are required to document those aspects of tests
considered of moderate or high complexity (Subpart K of 57 CFR 493),
including the need to establish and verify method-performance specifications.
These test aspects include accuracy, precision, reportable range, and
analytical detection limit.
Control Procedures
Under 1988 CLIA
regulations, control procedures should be in place that monitor instrument
stability and operator variance for all quantitative tests, and a daily
QC procedure should be performed that includes at least two samples
of different concentrations (normal and abnormal) to ensure the ongoing
validity of test results (§493.1218). For blood lead measurements, the
clinically relevant concentration range is 5 to 100 µg/dL of whole blood
although confirmed blood lead concentrations above 70 µg/dL are rare.
Most analytical methods are calibrated for a working range up to 60
µg/dL, with dilution required for specimens that are above the highest
calibration point. Most modern GFAAS methods have detection limits that
are significantly better than 5 µg/dL and can, therefore, report test
values of less than 5 µg/dL.
For daily QC monitoring,
a variety of materials are available for blood lead measurements. (See
Appendix C.2) We caution laboratories to verify independently each batch
of daily QC materials against a certified reference material before
they are used routinely. Any large deviation (1-2 µg/dL) from the stated
target value warrants further investigation.
The acceptable range
for routine daily QC should be no greater than that allowed for PT performance
samples (i.e., +4 µg/dL or +10%, whichever is greater).
Most laboratories experienced in analyzing blood lead levels should
have no difficulty in achieving a day-to-day precision that is better
than + 4 µg/dL at values of 10 µg/dL. Therefore, CDC’s Advisory
Committee on Childhood Lead Poisoning Prevention recommends that, for
investigative actions, laboratories set their internal QC limits to
+2 µg/dL or +10%, whichever is greater.
Although 1988 CLIA
regulations require a minimum of only two control concentrations (normal
and abnormal) for blood lead measurement, the Advisory Committee also
recommends that at least three levels of QC be employed at low (<
10 µg/dL); intermediate (25-30 µg/dL) and high (40-60 µg/dL) calibration
ranges for blood lead levels. The rationale for this recommendation
is that most daily QC problems with blood lead measurements arise with
the use of the high level control (> 40 µg/dL) because of sensitivity
drift. If the blood lead level of a single high-level control (50 µg/dL)
is out of the acceptable range, but the low-level control (10 µg/dL)
is within range, then all test results in the batch (e.g., tray carousel)
that are greater than the upper limit blood lead level of the low-level
control (i.e., > 14 µg/dL) would have to be rejected and the specimen(s)
reanalyzed. This reanalysis may create a problem if, for example, the
rejected results were capillary specimens, and there was not enough
sample remaining to perform a repeat analysis.
Under current regulatory
guidelines, test results above the upper limit of the lead level of
the low-level control can not be reported and the laboratory would request
that another specimen be obtained. Had an intermediate control (30 µg/dL)
been included in the run, the outcome might have been different. For
example, an intermediate control might have been within acceptable limits,
albeit with a low bias, but would have enabled the laboratory to report
the results of testing a capillary sample (e.g., results between 14-34
µg/dL) without having to obtain another specimen.
Another reason for
the tri-level approach is that with ASV analysis, when the electrode
plate begins to thin, the higher blood lead values begin to loose accuracy
(J. Chisolm, MD, Kennedy Krieger Institute, personal communication,
1993). Similarly, with GFAAS analysis, the most common problem is associated
with unacceptable errors with high-level control values (i.e., values
> 50 µg/dL).
Clinical Lead
Laboratory Approval and Accreditation
The CLIA regulations
of 1988 require that all clinical laboratories performing blood lead
tests participate successfully in an approved PT program. Currently,
five PT programs for proficiency in blood lead testing have been approved
by HCFA for CLIA purposes (Appendix C.2). Laboratories participating
in any one of three of these programs can also attain approval to perform
blood lead testing from OSHA.
In addition to proficiency-testing
requirements, many states require the mandatory reporting of blood lead
test results. Some states have or are moving toward electronic reporting
of all blood lead test results, whereas others require reporting elevated
levels only. As states move to update their definitions of elevated
blood lead tests results, many will require the mandatory reporting
of all tests results, and the most efficient means for doing so is via
electronic transmission. CDC’s Advisory Committee supports the concept
of electronic reporting of all blood lead test results by state since
doing so will facilitate CDC’s efforts in monitoring the incidence and
prevalence of lead poisoning nationally. For additional information
on electronic reporting of blood lead testing data, please contact your
state health department or CDC.
Turnaround Time
for Blood Lead test Results
Laboratory turnaround
time for blood lead testing will depend on several factors, including
the analytical method used, work-load fluctuations, and quality control
protocols used. Although most public health laboratories only operate
on a 9 a.m. to 5 p.m. weekday schedule, many private laboratories operate
around the clock, and therefore, may be able to reduce turnaround time
considerably. What is considered a reasonably turnaround time differ
for the different specimens that are analyzed. All laboratories recognize
the need to allocate a higher priority to those specimens considered
urgent (e.g., stat requests). However, capillary blood specimens,
which are considered for screening purposes only, should be analyzed
and reported to the requesting physician within 1 week of being received
in the laboratory.
Venous blood specimens,
which are more likely to be used for confirmatory follow-up purposes,
should receive a higher priority than that allocated to a screening
specimen. Confirmatory specimens should be analyzed and reported within
3 days of receipt in the laboratory. For stat requests, the laboratory
should be able to reduce turnaround time to 24 hours or less. A stat
request to determine a blood lead level from a capillary blood specimen
is inappropriate because such a request implies that the result will
be used to determine a course of medical management.
Reporting Blood
Lead Results
In the United States,
blood lead test results are usually reported in units of micrograms
per deciliter of whole blood (i.e., µg/dL or µg/100 mL. In most other
countries, the international system (SI) of units are preferred (i.e.,
micromoles per liter whole blood [µM or µmol/L]). To convert results
in µg/dL to µM, multiply the former by 0.048; for example, 10 µg/dL
- 0.48 µM.
A related aspect
of reporting a blood lead value is the rounding-off of the result. The
number of significant figures given is indicative of the precision of
the analytical method. The convention in analytical chemistry is to
report all the digits that are certain plus the first uncertain one.19
The precision of most routine blood lead methods developed before
the 1980s was only + 1-2 µg/dL or greater. Thus, routine blood
lead results were always rounded to the nearest integer for clinical
purposes.
Now that the blood
lead level of concern has been lowered from 25 µg/dL to 10 µg/dL of
blood lead, analytical methods with improved precision have been developed
for determining lower concentrations of lead in blood. For example,
modern GFAAS methods can attain a between-run precision of approximately
+ 0.3 µg/dL at low levels.8 Therefore, with repeated
analyses (n>3), blood lead test results at low levels could be reported
to the first decimal with some confidence if such reporting were required
(e.g., for research purposes). Of course, the laboratory would have
to validate within-run and between-run precision to justify reporting
results with such implied precision. For many routine blood lead testing
methods, however, such precision is not justified for a single analysis;
moreover, the clinical significance of such precision has not yet been
established. Therefore, routine blood lead test results should still
be rounded to the nearest integer.
Determining the
Level of Lead in Urine
In the past, the
primary purpose of determining lead levels in urine was to assess total
lead excretion over a fixed period, usually 8 hours, as a function of
the dose of calcium disodium ethylenedramine tetracetic acid (CaNa2EDTA)given.
Although using the lead mobilization or "provocation" test
has decreased in recent years and demand for testing urinary lead levels
has dropped, laboratories continue to receive specimens for analysis.
Tests for lead levels on non-timed or "spot" urine specimens
are considered to have little clinical value but are still used to monitor
workers occupationally exposed to organolead compounds. Some guidance
and recommendations for determining the level of lead in urine is provided
below.
Recommended Procedure
for Collecting Urine Samples
Collecting urine
for a provocation test is almost always performed on an inpatient basis,
and with very young children some additional difficulties occur in ensuring
that the entire 8-hour specimen is collected without contamination from
exogenous lead. A special commercially-available plastic pouch is taped
over the child’s genitalia to facilitate urine collection. However,
it is the responsibility of the laboratory providing the analysis to
ensure that all materials used to collect and transport urine specimens
are lead-free.
For an 8-hour urine
collection, the laboratory should provide a supply of primary urine
collection containers (1-L volume) with caps, preferably plastic. Containers
should be certified as lead-free, either by acid-washing them with 2%
(v/v) nitric acid for 24 hours, followed by washing them with deionized
water, or by selecting a small number from a batch, filling them with
2% nitric acid, and analyzing the leachate for lead. Either of these
procedures should prevent the use of contaminated containers.
Transporting
or Shipping Urine Specimens to Centralized Laboratories
If the urine lead
test is to be sent to a reference laboratory for analysis, then some
consideration should be given to packaging and shipping the specimen.
The laboratory performing the analysis should always be consulted because
it may have specific requirements or employ a courier service. Generally,
it is unnecessary and cumbersome to ship the entire urine specimen for
analysis since the cost will be higher and risks for leakage greater.
A 10-mL aliquot is usually sufficient for analysis, and this aliquot
can be easily transported in commercially available plastic syringes
designed specifically for transporting urine specimens.
There is no need
to refrigerate urine specimens during transit since they are quite stable
for several days. All urine collection materials should be provided
by the testing laboratory and certified as lead-free.
Using Additives
as Stabilizers
For determining
lead levels in urine, some laboratories have proposed stabilizing the
specimen with dilute nitric acid. However, the justification for acidification
is weak, and addition of acid provides an opportunity to contaminate
the specimen. Lead at high concentrations might precipitate out of solution
at a pH of 6-7, but this is hardly the case with urine, where the pH
is much lower. For this reason, it is unnecessary to add nitric acid
to the specimen.
Analytical Methods
for Determining Lead Levels in Urine
Most laboratories
use GFAAS to determine lead levels in urine; ASV may also be used for
the direct determination of urine lead levels, but analyzing urine by
ASV requires different treatment than analyzing blood by ASV. When using
GFAAS, the urine lead analysis can be more troublesome than that for
blood lead because of the high inorganic salt content in urine and the
lack of reliable urine reference materials with certified lead content
at clinically relevant concentrations.
Graphite Furnace
AAS
For some furnace
instruments, it may be possible to calibrate directly with aqueous lead
standards. If this is not possible, the best approach is to calibrate
with matrix-matched standards (i.e., with lead-spiked urine containing
nitric acid (2% v/v) and NH4H2PO4 modifier).
A donor should be sought who can provide urine samples containing little
or no detectable lead. Typically, the calibation range for urinary lead
is 0-60 µg/dL, as it is for blood lead. Because human urine can vary
considerably, a 1+9 dilution with modifier/acid will help offset any
potential interferences from other components present. As a rule of
thumb, the analysis should always be approached by referring to the
STPF concept.
Anodic Stripping
Voltammetry
ASV can be used
to determine urine lead levels, but several modifications are required
for this analysis to be successful. First, calibrators specific for
measuring urine lead levels should be used, and some laboratories recommend
that nickel be incorporated into the supporting electrolyte/decomplexing
solution to facilitate displacement of lead from EDTA (J. Chisolm, MD,
Kennedy Krieger Institute, personal communication, 1993).
Quality Assurance
and Quality Control
Few reliable reference
materials exist specifically for QA/QC of urine lead measurements. NIST
provides a bi-level Toxic Metals in Human Urine material (SRM 2670)
that is certified for lead at an "abnormal" level of 109 µg/L,
and a low level of 10 µg/L is given for informational purposes. However,
this material has only limited application for clinical purposes, where
typical urine concentrations range from 50 to 1000 µg/L. Using a 1+9
sample dilution, a linear calibration range up to 60 µg/L gives directly
reportable concentrations of up to 600 µg/L. Clearly there is a need
for reliably validated urine lead reference materials at concentrations
between 100 and 600 µg/L. A list of urine control-material sources is
provided in Appendix C.2.
Performance Criteria
for Analyzing Lead in Urine
There are no established
performance criteria for analyzing urine lead levels. Currently, no
proficiency testing exists specifically for urine lead because the test
is so rarely performed. Nontheless, it should be possible for analytical
laboratories to achieve a level of performance of + 60 µg/L at
concentrations of <400 µg/L and + 15% at concentrations above
400 µg/L. In fact, these were the performance standards expected for
blood lead determinations before 1991, and they reflect the actual performance
of a group of laboratories performing urine lead determinations using
GFAAS, ASV, and MIBK-FAAS.20
Erythrocyte Protoporphyrin
Test
The erythrocyte
protoporphyrin (EP) test was used for many years throughout the United
States as a screening test to identify children exposed to lead. In
the 1991 edition of Preventing Lead Poisoning Prevention in Young
Children, CDC recommended that EP no longer be used as a screening
test to detect lead-exposed children.21 The justification
for this recommendation was that the results of numerous studies showed
the poor diagnostic sensitivity of EP for detecting blood lead levels
at
10 µg/dL, and even
at 25 µg/dL, coupled with an equally poor
specificity.22,23
However, EP remains
a valuable test in the medical management and follow-up care of children
with confirmed elevated blood lead levels and as a screening test for
iron deficiency. Blood lead-EP pairs are particularly useful in following
long-term trends in lead absorption and in evaluating the question of
internal redistribution of lead after chelation therapy.
Specimen Collection
Either capillary
or venous blood may be used for the EP test; however, blood specimens
should be protected from prolonged exposure to light. For example, evacuated
glass tubes can be wrapped in aluminum foil. The preferred anticoagulant
for this test is EDTA, although heparinized blood may be also used.
Instructions on capillary collection are given in Appendix C.2.
Analytical Methods
for Determining EP
There are two principal
methods for determining EP, acid-extraction and hematofluorometry. The
extraction method is generally accepted as the reference method for
EP and involves extracting protoporphyrin and other heme components
from whole blood into a mixture of ethyl acetate and acetic acid and
back-extracting the protoporphyrin into dilute hydrochloric acid. Quantitation
is performed using a spectrofluorometer calibrated with protoporphyrin
IX standards. A complete description of the extraction method is beyond
the scope of this document; refer to the National Committee for Clinical
Laboratory Standards’ (NCCLS) document CP42-P on Erythrocyte Proroporphyrin
testing.24
Hematofluorometry
(HF) uses a small portable fluorometer dedicated specifically to measuring
EP directly in whole blood as the zinc chelate, zinc protoporphyrin
(ZPP). This instrument was once widely used in public health programs
to screen children for lead exposure. Again, a complete description
of the use and issues related to hematofluorometry is beyond the scope
of this document; refer to the NCCLS document on EP testing mentioned
above.
Millimolar Absorptivity
(mÎ) of Protoporphyrin IX
One issue yet to
be completely resolved is continued widespread use of an incorrect millimolar
absorptivity (mÎ) value (241 L cm-1 mmol-1)for
protoporphyrin IX calibration standards. Since most hematofluorometers
trace calibration back to the reference extraction method, the mÎ
issue affects hematofluorometry as well. Historically, the mÎ
value was thought to be 241L cm-1 mmol-1, but
this value has been recently shown to be incorrect and the true value
is 297 L cm-1 mmol-1.25 The impact
of using the correct mÎ value is a 19% decrease in all EP test
results, including reference ranges. However, individual laboratories
in the United States should not initiate any changes in calibration
procedure unilaterally but should follow directives from the Health
Resources and Services Administration (HRSA)-sponsored PT program for
EP, operated by the Wisconsin State Laboratory of Hygiene, and from
state PT programs for EP because such changes will have an impact on
hematofluorometer calibration, target values for standards and controls,
and reference ranges for EP.
Reference Ranges
for EP
In the 1985 edition
of Preventing Lead Poisoning in Young Children,26 CDC
adopted an EP value of 35 µg/dL of whole blood as the upper limit of
normal for screening children for lead exposure. This level corresponded
to a blood lead level of 25 µg/dL. A simple adjustment of that value,
correcting for the mÎ error, results in a lower value of 28 µg/dL,
which was adopted in the 1991 edition of Preventing Lead Poisoning
in Young Children.21 The NCCLS document on EP testing
reports the correct upper limit of normal for EP, using mÎ of
297 L cm-1 mmol-1, to be 30 µg/dL, partly on the
basis of reanalysis of the NHANES II data set, for which EP values were
originally determined using an mÎ value of 297 L cm-1
mmol-124. Thus, the CDC Advisory Committee, recognizing that
EP has been plagued by historical inaccuracies in the analysis, recommends
that the upper limit of normal for an EP test result is 30 µg/dL of
whole blood. However, the Advisory Committee also reiterates that individual
laboratories should follow guidance from PT program officials on the
timing and manner in which this change in calibration should occur.
Standards and
Control Materials for EP Testing
The availability
of standards and control materials for use in determining EP is limited.
Pure protoporphyrin IX standards are available from only one source
(Porphyrin Products, Logan, Utah). For extraction methods, lyophilized
whole blood control materials are available from several sources and
have been used as either control specimens or as secondary standards.
Frozen whole-blood materials (human, bovine, or goat), if properly stored
and protected from light, may also be used but are not commercially
available. Hemolyzed or reconstituted lyophilized blood cannot be used
with hematofluorometry because it requires the presence of intact red
blood cells for correct quantitation. Since HF calibration is usually
provided by the manufacturer, daily quality control is limited to testing
liquid blood materials manufactured specifically for the hemotofluorometer
(Appendix C.2).
Proficiency Testing
Programs for EP
A limited number
of PT programs are available for EP. They include the HRSA-sponsored
program operated by the Wisconsin State Laboratory of Hygiene that is
open to anyone and several state-based PT programs (e.g., in New York,
Pennsylvania, and New Jersey), where participation is required for in-state
permit purposes. (See Appendix C.2). Although participation in a PT
program for EP is not required by HCFA under the 1988 CLIA regulations,
the Advisory Committee strongly recommends that all laboratories performing
the analysis for EP participate successfully in such a program.
Miscellaneous
Tests for Lead in Biological Fluids or Tissues
Lead Levels can
be determined in a number of different tissues (teeth, hair, nails)
and body fluids (breast milk, sweat, ). However, the clinical utility
of such analyses is doubtful, and some tissues (e.g., hair, nails),
may be so grossly comprised by contamination errors as to make their
analysis totally unreliable. Tests to determine lead levels in teeth
have proven useful in clinical research studies, 26,27 because
they reflect cumulative lead exposure but are less valuable as routine
clinical tests. Teeth may not always be available for specimens and
may be compromised by inappropriate storage. The analysis is complicated
by the fact that the tooth specimen must be digested under clean conditions
before its lead content can be measured.
Digestion techniques
are not routinely practiced by most clinical laboratories. No reference
materials or controls are available to validate such analytical procedures
and, given the regulatory requirements of CLIA, many laboratories capable
of performing the analysis are reluctant to offer it because of such
obvious difficulties. The same difficulties hold true for tests for
lead levels in milk, hair, and nails.
*A selected list
of methods of measuring blood lead levels and obtaining information
about instrument manufacturers and other information pertinent to the
clinical laboratory is available from CDC’s National Center for Environmental
Health, Division of Environmental Health Laboratory Sciences, Nutritional
Chemistry Branch, MS F18, 4770 Buford Highway, N.E., Atlanta, Georgia
30341-3274, telephone (770) 488-4452.
References
1. Delves H.T. Blood
collection for screening children for exposure to lead [[letter]. Clin
Chem 1996;42:983-4.
2. Wang ST, Peter
F. The stability of human blood lead in storage. J Anal Toxicol 1985;9:85-8.
3. Hessel DW. A
simple and rapid quantitative determination of lead in blood. Atomic
Absorption Newsletter 1968 7:55-6.
4. Delves HT. A
micro-sampling method for the rapid determination of lead in blood by
atomic absorption spectrophotometry. Analyst 1970;95:431-8.
5. Pruszkowska E,
Carnick GR, Slavin W. Blood lead determination with the platform-furnace
technique. Atomic Spectroscopy 1983 4:59-61
6. Shuttler IL,
Delves HT. Determination of lead in blood by atomic absorption spectrometry
with electrothermal atomisation. Analyst 1986;111:651-6.
7. Miller DT, Paschal
DC, Gunter EW, Stroud PE, D’Angelo J. Determination of lead in blood
using electrothermal atomization atomic-absorption spectrometry with
a L’vov platform and matrix modifier. Analyst 1987;112:1701-4.
8. Parsons PJ, Slavin
W. A rapid Zeeman graphite furnace atomic absorption spectrometric method
for the determination of lead in blood. Spectrochimica Acta, Part B
1993; 48B:925-939.
9. Subramanian KS.
Determination of lead in blood by graphite furnace atomic absorption
spectrometry--a critique. Science of the Total Environment 1989;89:237-350.
10. Qiao H, Parsons
PJ, Slavin W. Transferability of a simple method for blood lead determination
using furnace atomic absorption spectrometry and continuum background
correction. Clin Chem 1995;41:1451-4.
11. Flajnik CE,
Shrader D. Determining lead in blood: evaluating deuterium and Zeeman
background correction. American Clinical Laboratory 1994;13(9):45-6.
12. Helfrich JP.
A workstation-based quick furnace method for the determination of lead
in whole blood. American Clinical Laboratory 1994;13(4):14-5.
13. Carnrick GR,
Slavin W. Modern graphic furnace AAS: part 1 and part 2. American Laboratory
1988;44B:188-97.
14. Slavin W. Graphite
furnace AAS for biological materials. Science of the Total Environment
1988;71:17-35.
15. Campbell MJ,
Delves HT. Accurate and precise determination of lead isotope ratios
in clinical and environmental samples using inductively coupled plasma
source mass spectrometry. J Anal Atomic Spectrom 1989;4:235-6.
16. Roda SM, Greenland
RD, Bornschein RL, Hammond PB. Anodic stripping voltammetry procedure
modified for improved accuracy of blood lead analysis. Clin Chem 1988;34:563-7.
17. Jagner D, Josefson
M, Westerlund S, Aren K. Simultaneous determination of cadmium and lead
in whole blood and serum by computerized potentiometric stripping analysis.
Anal Chem 1981;53:1406-10.
18. Wang J, Tian
B. Screen-printed stripping voltammetric/potentiometric electrodes for
decentralized testing of trace lead. Anal Chem 1992;15:1706-9.
19. Miller JC, Miller
JN. Statistics for analytical chemistry. Second edition. Chichester,
United Kingdom: Ellis Horwood Ltd., 1988.
20. Parsons PJ,
Jackson KW, Hughes M, Chace SW, Jenkins S. The influence of chemical
modifiers on the atomization of lead from human urine by electrothermal
atomic absorption spectrometry [abstract]. Clin Chem 1990;36:970-3.
21. Centers for
Disease Control. Preventing lead poisoning in young children: a statement
by the Centers for Disease Control. Atlanta, GA: Department of Health
and Human Services, 1991.
22. McElvaine MD,
Orbach HG, Binder S, Blanksma LA, Maes EF, Krieg RM. Evaluation of the
erythrocyte protoporphyrin test as a screen for elevated blood lead
levels, J Pediatr 1991;119:548-50.
23. Parsons PJ,
Reilly AA, Hussain A. Observational study of erythrocyte protoporphyrin
screening test for detecting low lead exposure in children: impact of
lowering the blood lead action threshold. Clin Chem 1990;37:216-25.
24. National Committee
for Clinical Laboratory Standards (NCCLS). Erythrocyte protoporphyrin:
proposed guideline (C42-P). NCCLS document C42-P. Wayne, PA: NCCLS,
1995.
25. Gunter EW, Turner
WE, Huff, DL. Investigation of protoporphyrin IX standard materials
used in acid-extraction methods and a proposed correction for the millimolar
absorptivity of protoporphyrin IX. Clin Chem 1989;35:1601-8.
26. Centers for
Disease Control. Preventing lead poisoning in young children: a statement
by the Centers for Disease Control. Atlanta, GA: Department of Health
and Human Services, 1985.
27. Rabinowitz MB,
Leviton A, Bellinger DC. Blood lead-tooth lead relationship among Boston
children. Bull Environ Contam Toxicol 1989; 43:485-92.
28. Rabinowitz MB,
Leviton A, Bellinger D. Relationships between serial blood leadlevels
and exfoilated tooth dentin lead levels: models of tooth lead kinetics.
Calcified Tissue International 1993;53:338-41.
This paper is
printed with permission from the US Centers for Disease Control and
Prevention. Source: CDC Technical Papers.
TOP